Autosampler Configuration for Node Control (For Old & New Versions)

Attributed iFixit repair guide for Open-Storm Hardware.

iFixit Library Open-Storm Hardware Moderate 5 minutes Guide 123959
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Step 1

Starting the configuration process.

  • 1. Turn on the autosampler. Screen will read "...STANDBY..."
  • 2. Press "ENTER/PROGRAM" button. Use arrows to highlight "CONFIGURE". Press "ENTER/PROGRAM" button
  • This will allow you to use the arrows to shuffle through the different configuration options. To select the one you want, press the "ENTER/PROGRAM" button.
  • 3. Use arrows to find "EVENT MARK". Press the "ENTER/PROGRAM" button.
Step 2

Configuring "event mark".

  • 4. Use arrow to select "CONTINUOUS SIGNAL". Press "ENTER/PROGRAM" button.
  • 5. Use arrow to select "PUMP CYCLE". Press "ENTER/PROGRAM" button.
  • It is critical this setting is configured in order for the sensor node to operate the autosampler!
  • 6. Use arrows to find "EXIT CONFIGURATION." Press "ENTER/PROGRAM" button. This will bring up the screen "...STANDBY...