Siemens Maquet Servo-i Pre-Use Check

The Pre-Use Check includes tests and measuremen...

iFixit Library Siemens-Maquet Servo-i Moderate No estimate Guide 132176
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

The Pre-Use Check includes tests and measurements of: internal technical functionalityinternal leakagepressure transducersO2 cell / O2 sensorflow transducerssafety valvebattery modulespatient breathing system leakagecircuit compliance WARNINGS Always perform a Pre-Use Check before connecting the ventilator to a patient.The separate Patient Circuit Test that can be performed in Standby mode does not replace the Pre-Use Check.Do not connect the ventilator to a patient while a malfunction persists.Do not lift the expiratory cassette while the ventilator is in operation; instead, lift the cassette while in Standby mode. Important: If you change the breathing circuit after calculating the circuit compliance compensation factor, perform a new Pre-Use Check.

The iFixit procedure is temporarily unavailable through the API.