Zoll Impact Uni-Vent 731 Power Knob Kit Replacement

This information can be found on page 29 of the...

iFixit Library Zoll Impact Uni-Vent 731 Moderate 20 minutes Guide 133652
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

This information can be found on page 29 of the Zoll Impact Uni-Vent 731 Ventilator Guide on the iFixit page for the ventilator. This is a simple repair that can be done on this model of Zoll ventilator and should be done whenever there is damage to the power knob.

Step 1

Repair step

  • Loosen but do not remove the(4) 8-32 X 3 screws on the outer air intake
  • Pg. 29
Step 2

Repair step

  • Loosen and remove the (2) 10-32 Keps nuts and the (4) 6-32 X 2 screws. Remove the back case by lifting from the ventilator
  • Pg. 29
Step 3

Repair step

  • Remove the (2) 4-40 X ¼ screws on the Dovetail Mounting Bracket and disconnect the ribbon cable on the PIM PCB by simultaneously applying pressure on the two locking “ears”
  • Pg. 29
Step 4

Repair step

  • Flip the ventilator over and remove the front case assembly by lifting it straight up away from the ventilator module
  • Pg. 29
Step 5

Repair step

  • Remove the bezel by loosening and removing the (7) 4-40 X ¼ screws that hold the bezel to the front case
  • Pg. 29
Step 6

Repair step

  • On the membrane panel assembly, loosen the 6-32 X ¼ set screw using a 1/16 Ball Hex Driver (located in the power knob) and lift the damaged knob from the membrane panel
  • Pg. 30
Step 7

Repair step

  • Place the replacement knob on the membrane panel switch -align with flat on switch –insert the included 6-32 X ¼ set screw and tighten
  • Pg. 30
Step 8

Repair step

  • Install the new bezel by tightening the included (7) 4-40 X ¼ screws
  • Pg. 30
Step 9

Repair step

  • Place the front case assembly over the vent module and tighten the (2) 4-40 X ¼ screws unto the dovetail mounting bracket. Reconnect the ribbon cable unto the PIM PCB. Make sure the two locking “ears” lock into position
  • Pg. 30
Step 10

Repair step

  • Attach the back case to the vent module and align cover with handle, air intake housing and dovetail mounting studs. Insert and tighten the included (2) 10-32 Keps nuts and the (4) 6-32 X 2” screws
  • Pg. 30
Step 11

Repair step

  • Tighten the (4) 8-32 X 3 screws on the outer air intake
  • Pg. 30
Step 12

Repair step

  • Perform HiPot Testing then perform Calibration and Functional Test using the RCS
  • Pg. 30