Zoll Impact Uni-Vent 731 PIM Board Kit Replacement

The following guide outlines the process to...

iFixit Library Zoll Impact Uni-Vent 731 Moderate 45 minutes Guide 133656
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

The following guide outlines the process to repair the PIM Board Kit of the Zoll Ventilator. This guide is useful in the case that the PIM Board fails to work due to a power failure or missing battery. This guide allows one to find the steps to repair this guide quickly and efficiently. For more information on replacing the PIM Board, please refer to page 46 of the service manual.

Step 1

Repair step

  • Loosen but do not remove the (4) 8-32 X 3 screws on the outer air intake
Step 2

Repair step

  • Loosen and remove the (2) 10-32 Keps nuts and the (4) 6-32 X 2 screws. Remove the back case by lifting from the ventilator.
Step 3

Repair step

  • Remove the (2) 4-40 X ¼ screws on the Dovetail Mounting Bracket and disconnect the ribbon cable on the PIM PCB by simultaneously applying pressure on the two locking “ears”.
Step 4

Repair step

  • Flip the ventilator over and remove the front case assembly by lifting it straight up away from the ventilator module.
Step 5

Repair step

  • Remove the battery compartment cover by unscrewing the (4) 6-32 X 5/16 screws.
  • Remove the battery by unscrewing the (4) 6-32 X 2 ¼ screws and detaching the plug from its locking latch.
Step 6

Repair step

  • Disconnect the Power Input cable from the PIM PCB by pressing on the locking latch and pulling the cable straight up from the connector.
Step 7

Repair step

  • Loosen and remove the(5) 4-40 X 5/16 screws holding the PIM PCB to the ventilator module. Position SPO2 Insulator out of the way (Do Not Fold) and insert screwdriver through the holes to aid in removing screws.
Step 8

Repair step

  • Lift the defective PIM Board out of the ventilator module.
Step 9

Repair step

  • Place the new PIM Board Unto the vent module (make sure male header pins are inserted correctly into the mating header.
Step 10

Repair step

  • Secure the PIM Board with the (5) 4-40 x 5/16 screws.
Step 11

Repair step

  • Re-assemble the battery by connecting its cable to the connector (pull on cable to insure it is locked in place) then tightening the (4) 6-32 X 2 ¼ screws.
  • Re-assemble the battery compartment cover by tightening the (4) 6-32 X 5/16 screws.
Step 12

Repair step

  • Place the front case assembly over the vent module and tighten the (2) 4-40 X ¼ screws unto the dovetail mounting bracket. Reconnect the ribbon cable unto the PIM PCB. Make sure the two locking “ears” lock into position.
Step 13

Repair step

  • Attach the back case to the vent module and align cover with handle, air intake housing and dovetail mounting studs. Insert and tighten the included (2) 10- 32 Keps nuts and the (4) 6-32 X 2” screws.
Step 14

Repair step

  • Tighten the (4) 8-32 X 3 screws on the outer air intake.
Step 15

Repair step

  • Perform HiPot Testing then perform Calibration and Functional Test using the RCS.