Weinmann VENTImotion and VENTIlogic Power Board Replacement

Follow this replacement guide to replace your...

iFixit Library Weinmann VENTImotion and VENTIlogic Moderate 15 minutes Guide 133741
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

Follow this replacement guide to replace your power board. You will need to replace the power board if there is no running noise and displays will not show up or, when in automatic mode, the device cannot be switched on by a breath being taken in. For this guide, you will need a 3mm Allen Key. Be cautious when disconnecting items so that you no not accidentally break them and remember where you pulled them out so that reassembly is easy. The steps are not difficult but should be performed carefully. For more information on replacing the power board, please refer to page 34 in the service manual.

Step 1

Repair step

  • Open the Device by following these steps:
  • Open the device only with the power supply plug disconnected.
  • Place the device on a non-slip surface with the top facing downwards.
  • Remove the filter cassette from the top part of the housing
  • Undo and remove the 6 screws (34)
  • Open the bottom part of the housing (35) away to the side.
  • Undo the connecting cables for the humidifier (42) and the alarm (43)
  • You can now put the bottom part of the housing (35) to one side.
Step 2

Repair step

  • Disconnect the power board connecting cable (45) from the power board (41).
Step 3

Repair step

  • Disconnect the Motor Cable
Step 4

Repair step

  • Disconnect the O2 valve/power board connecting cable (70) from the power board.
Step 5

Repair step

  • Disconnect all connecting cables from the control board.
Step 6

Repair step

  • Remove the power board from the top part of the housing.
Step 7

Repair step

  • Take the rechargeable battery connecting cable (48) out of the groove and disconnect it from the power board (41).
Step 8

Repair step

  • Disconnect the cables of the old power board and plug them into the appropriate slots of the new power board 41:
  • –power board/control board connecting cable (46)
  • –rechargeable battery connecting cable (48)
  • –earth cable (47)
Step 9

Repair step

  • Connect the power board/mains supply unit connecting cable (45) to the new power board.
Step 10

Repair step

  • Position the new power board in the top part of the housing.
Step 11

Repair step

  • Restore the connections to the control board.