Weinmann VENTImotion and VENTIlogic Control Board Replacement

Follow this replacement guide to replace your...

iFixit Library Weinmann VENTImotion and VENTIlogic Moderate 30 minutes Guide 133773
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

Follow this replacement guide to replace your control board. You will need a 3mm Allen Key and a size 1 Phillips screwdriver before starting on this venture. There are more than a few steps but this guide also overlays other guides. Such as “opening the device”, “Replacing the Display” and “Changing the Battery of the Control Board”. I would set aside time and patience for this task. You can also check out the device’s troubleshooting page VENTImotion and VENTIlogic Ventilator Troubleshooting For more information on replacing the control board, please refer to page 39 in the service manual.

Step 1

Repair step

  • Open the device
  • 1) Place the device on a non-slip surface with the top facing downwards.
  • 2) Remove the filter cassette from the top part of the housing.
  • 3) Undo and remove the 6 screws (34)
Step 2

Repair step

  • Remove the power board (see Weinmann VENTImotion and VENTIlogic Power Board Replacement)
Step 3

Repair step

  • Disconnect the following
  • - encoder connector (82)
  • -connectors of interfaces (84) and (85) of the control board.
Step 4

Repair step

  • Disconnect the ribbon cable of the display and the fascia film:
  • Pull the latch out until you feel resistance. You can then pull out the ribbon cable.
  • If the latch has not been opened correctly, the ribbon cable may be damaged when it is pulled out.
Step 5

Repair step

  • Pull flow measurement hoses (56) off the flow sensor.
Step 6

Repair step

  • Pull pressure measurement hoses (57) off the pressure sensors.
Step 7

Repair step

  • Insert a new battery (52) in the appropriate holder of the new control board
  • Screw on spring (54)
Step 8

Repair step

  • Put the control board in the top part of the housing and attach it with the four screws (51)
Step 9

Repair step

  • Reattach the pressure measurement hoses (57) to the pressure sensors.
  • Connect the flow sensor marked Hi to the hose connector marked Hi at the device outlet.
Step 10

Repair step

  • Attach the flow measurement hoses (56) to the flow sensor
  • Connect the connection of the flow sensor marked Lo to the hose connector marked Lo at the device outlet.
Step 11

Repair step

  • To prevent the hoses from kinking, position flow measurement hoses (56) and pressure measurement hoses (57) as shown.
Step 12

Repair step

  • Connect the ribbon cable for the fascia film and the display to the control board again:
  • –pull the latch upwards.
  • –push the ribbon cable into the connector.
  • –press the latch back down
Step 13

Repair step

  • Plug the connectors of the interfaces and the rotary knob into the appropriate slots of the control board.
Step 14

Repair step

  • Refit the box and filter holder.
  • Connect the box and the filter holder to one another by reconnecting the intake side hose (73) from both parts.
  • Refit box (58) and filter holder (37) back on the top part of the housing together.
  • Connect the pressure-side hose (74) from the box.
Step 15

Repair step

  • Refit the power board Weinmann VENTImotion and VENTIlogic Power Board Replacement
Step 16

Repair step

  • Close the Device: Hold the bottom part of the housing (35) up to the side of the top part of the housing (36)
  • Plug the connecting cables for the humidifier (42) and the alarm (43) onto the relevant connectors.
  • Check that all hoses and cables are plugged on firmly.
  • Place the bottom part of the housing (35) on the top part of the housing (36). Ensure that no cables or hoses are trapped or bent.
  • Now screw the top part of the housing tight using the 6 screws (34).
  • Then turn the device back over
Step 17

Repair step

  • Check contrast and adjust if necessary .
Step 18

Repair step

  • Reset the date and time.
  • Set the date as follows: Year, Month, Day
Step 19

Repair step

  • Test the device.
  • Page 13 of the VENTImotion & VENTIlogic Service and Repair Instructions will go into more detail.