How to Open and Close Housing (SOMNOcomfort and SOMNObalance Ventilator)

In order to conduct repairs or maintenance, the...

iFixit Library SOMNOcomfort and SOMNObalance Ventilator Moderate 5 - 10 minutes Guide 133792
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

In order to conduct repairs or maintenance, the SOMNOcomfort and SOMNObalance ventilator must be opened and closed. This guide goes through each step necessary to access the internal components of the SOMNOcomfort and SOMNObalance ventilators. The page numbers included stem from the service repair guide (linked on the device page).

Step 1

Repair step

  • Remove the covering cap. (pg. 23)
Step 2

Repair step

  • Place the unit upside down on a non-slip surface and remove the six screws. (pg. 23)
Step 3

Repair step

  • Remove the lower part of the housing with a swiveling motion away to the side. Note the SOMNOcomfort series uses a shorter circuit board than the balance series ventilator. (pg. 23)
Step 4

Repair step

  • Remove the connector (X1) from the circuit board.
  • Pull the connecting cable (64) from its mount.
  • Now set aside the lower part of the housing. (pg. 23)
Step 5

Repair step

  • Next, how to close the compartment and reverse the process.
  • Hold the lower part of the housing to the upper part at an angle
  • Attach the connector (X1) to the main circuit board. When doing so the connector’s nub has to point towards the lock. (pg. 24)
Step 6

Repair step

  • Properly align the main circuit board inside the unit.
  • Attach the connecting cable (64) to its mount (pg. 24)
Step 7

Repair step

  • Place the lower part of the housing on the upper part (47). Please make sure the main circuit board (44) is aligned properly, and that the connecting cable (64) is not being squeezed anywhere. (pg. 24)
Step 8

Repair step

  • Now reattach the upper part of the housing by tightening the six screws. (pg. 24)
Step 9

Repair step

  • Insert the covering cap. (pg. 24)