Puritan Bennett 840 Oxygen Sensor Replacement

Use this guide to replace the oxygen sensor on...

iFixit Library Puritan Bennett 840 Moderate 5 - 10 minutes Guide 133802
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

Use this guide to replace the oxygen sensor on a Puritan Bennett 840 ventilator. The oxygen sensor should be replaced every two years for preventative maintenance. If being used very frequently, or the oxygen sensor starts to fail, replace as often as necessary. When servicing the ventilator, be sure to familiarize yourself with, and adhere to all posted and stated safety warning and caution labels on the ventilator and its components, and on any service equipment and materials used. Failure to adhere to such warnings and cautions at all times may result in injury or property damage. Puritan-Bennett recommends that customers and technical services personnel follow local governing ordinances and recycling instructions regarding disposal or recycling of battery or other device components. For more information on replacing the oxygen sensor (OS), please refer to page 8-61 (section 8.15.8.11) in the service manual.

Step 1

Repair step

  • Locate the flexible oxygen sensor access cover on the top edge of the cabinet and firmly push the center of the lower flap of the access cover until the lower flap is dislodged from the cabinet.
Step 2

Repair step

  • Pinch the bottom and top flaps of the access cover firmly together and pull the access cover away from the cabinet to remove.
Step 3

Repair step

  • Press the small white release tab away from the sensor cable connector to release the sensor cable connector.
  • While continually pressing the release tab, gently pull the connector from the oxygen sensor.
Step 4

Repair step

  • Unscrew (counter-clockwise) and remove the oxygen sensor.
Step 5

Repair step

  • Open the replacement oxygen sensor and O-ring. Make sure the O-ring is at the threaded base of the oxygen sensor, and screw (clockwise) until snug.