Newport e500 Exhalation Module Disassembly

If you need to access important parts...

iFixit Library Newport e500 Very easy 2 minutes Guide 133820
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

If you need to access important parts associated with the Newort e500, removal of the exhalation module is an integral step. This step serves as a common pre-requisite to completing other guides. Make sure you are careful with the outlet boot as it is fragile. For more information refer to page 42 of the service manual.

Step 1

Repair step

  • Pull up and remove the flexible rubber exhalation outlet boot (push fit).
Step 2

Repair step

  • Disconnect the tubes from both sides of the flow sensor and the two tubes from the exhalation valve.
Step 3

Repair step

  • Pull the exhalation module straight up to remove it from the GDU.