Hamilton Galileo Ventilator Power Supply Replacement

Utilize this guide to replace the power supply....

iFixit Library Hamilton Galileo Ventilator Easy 8 minutes Guide 133862
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

Utilize this guide to replace the power supply. This guide requires access to the column, so you will need to view the “Opening the Column” guide as well. For more information on replacing the power supply, please refer to page 11-179 in the full service manual.

Step 1

Repair step

  • Use a grounding cable and wrist band connected to the GALILEO or other work surface when you are working on the GALILEO.
Step 2

Repair step

  • Remove from GALILEO: External mains power supply, Air and oxygen supplies, Patient breathing circuit (for convenience only), Flow Sensor tubes (for convenience only).
Step 3

Repair step

  • Loosen the rear panel of the column by removing the six hexagonal-drive (Allen) screws that hold it in place.
Step 4

Repair step

  • Without removing any electrical connections, open the left side of the column as wide as practical.
  • This exposes: GALILEO’s internal power supply, The battery fuse, The diagnostic LEDs on the power supply’s main circuit board, and the back up batteries.
Step 5

Repair step

  • Make sure no electrical connections are touching the column.
  • Make sure all green/brown grounding (earthing) cables are securely in place.
Step 6

Repair step

  • Remove all cables connecting power supply to GALILEO.
  • To remove the main power connector, squeeze on the two points indicated in the second picture.
Step 7

Repair step

  • Remove the rear cover and place it on a suitable surface.
Step 8

Repair step

  • Unscrew the retaining strip securing the power supply in its housing.
Step 9

Repair step

  • Slide the power supply out from its housing.
Step 10

Repair step

  • To install the new power supply, simply do steps 5-2 in reverse.
  • Make sure the security clamp (shown in picture) is tight before reinstalling power supply.
Step 11

Repair step

  • Perform the internal voltage test linked here: page 7-5 of Section 7.3, Checking voltages.
  • If this test is not successful, it is likely that there is a loose connection somewhere between the internal power supply and the printed circuit boards. Check the wires to make sure they are securely fastened.
  • When the above test is successful, continue with the tests in the following sections:
  • 1) Section7, Checking hardware, voltages, and interface
  • 2) Either: Section 9, Running Original Upgrade 1 test software OR Section 10, Running Upgrade 2 test software