Newport e500 Internal Battery Disassembly

This guide will cover the removal and...

iFixit Library Newport e500 Easy 10 minutes Guide 133966
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

This guide will cover the removal and replacement of the internal battery situated within the Newport e500. In order to perform this operation, a screwdriver and torque wrench will be required. Make sure the battery terminals do not come into contact with any other metal object when handled. For more information refer to page 68 of the service manual.

Step 1

Repair step

  • Place device so that the top cover is facing you.
  • Open the top cover.
Step 2

Repair step

  • Remove the four hex nuts and lock washers (two at each hinge) that secure the cover to the hinges on the left side of the pneumatic box.
Step 3

Repair step

  • Remove the cover and set aside. Retain hardware for re-assembly.
Step 4

Repair step

  • Pull up and remove the flexible rubber exhalation outlet boot (push fit).
Step 5

Repair step

  • Disconnect the tubes from both sides of the flow sensor and the two tubes from the exhalation valve.
Step 6

Repair step

  • Pull the exhalation module straight up to remove it from the GDU.
Step 7

Repair step

  • Remove the two nuts that hold the battery cover to the electronics box assembly and remove the cover.
Step 8

Repair step

  • Disconnect the black and red wires from the battery terminals.
Step 9

Repair step

  • Pull the battery from the electronics box assembly.