FIXONIC
SellSavedMy Fixonic
Official iFixit API visual reference

Newport e500 Pneumatic Panel Disassembly

This guide will show you how to remove the...

iFixit Library Newport e500 Moderate 15 - 20 minutes Guide 134016
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

This guide will show you how to remove the pneumatic panel for further disassembly and replacement. For some helpful tips, go to page 62 in the user manual.

Step 1

Repair step

  • Place device so that the top cover is facing you.
  • Open the top cover.
Step 2

Repair step

  • Remove the four hex nuts and lock washers (two at each hinge) that secure the cover to the hinges on the left side of the pneumatic box.
Step 3

Repair step

  • Remove the cover and set aside. Retain hardware for re-assembly.
Step 4

Repair step

  • Pull up and remove the flexible rubber exhalation outlet boot (push fit).
Step 5

Repair step

  • Disconnect the tubes from both sides of the flow sensor and the two tubes from the exhalation valve.
Step 6

Repair step

  • Pull the exhalation module straight up to remove it from the GDU.
Step 7

Repair step

  • Remove the 2 screws and washers securing the manifold/mixer assembly
Step 8

Repair step

  • Slide the manifold/mixer assembly forward to disengage the flow sensors
Step 9

Repair step

  • Gently pull each flow sensor straight out from the servo valve, (the flow sensor push-fits into the servo valve port).
Step 10

Repair step

  • Disconnect the flow sensor cable from its 10 pin connector at the underside of the flow sensor PCB and lift out the flow sensor.
Step 11

Repair step

  • Remove the screw and washer that secure the exhalation manifold support to the pneumatic panel.
Step 12

Repair step

  • Pull up on the exhalation valve guide block to remove it from the pneumatic box.
Step 13

Repair step

  • Remove the standoff and five screws and washers that secure the pneumatic panel to the pneumatic box.
Step 14

Repair step

  • Lift the pneumatic panel out of the box, and from the underside, detach all tubes and the heater wires that connect to the analog PCB.