Philips Respironics V60 Top Cover Replacement

If there has been damage done to the top cover...

iFixit Library Philips Respironics V60 Easy 25 minutes Guide 134021
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

If there has been damage done to the top cover of the Philips Respironics V60 ventilator, this guide will help you replace it. The top cover just simply serves as a cover for the inside of the ventilator, making sure that the parts of the ventilator within are protected. Damage to this could also damage the inside of the ventilator so it would be good to replace/repair it as quickly as possible.

Step 1

shut down the ventilator

  • Press and release the ON/Shutdown key, opening the Shutdown window in the screen.
  • Select Ventilator Shutdown, which will shut down the ventilator.
  • Improper shutdown may cause a Power has been restored message the next time the ventilator is turned on.
Step 2

disconnect from power

  • Disconnect the ventilator from the AC power.
Step 3

Repair step

  • Unplug the ventilator
Step 4

Repair step

  • Disconnect the internal battery from the power harness connector.
Step 5

Repair step

  • Turn the captive D-ring fastener on the bottom of the panel counter-clockwise one-quarter turn and release.
  • Then remove the right side panel.
Step 6

Repair step

  • Remove the 4 screws using a screwdriver.
  • Remove the top cover.