Before you begin
The nozzles are part of the Maintenance Kit 5000 hours that should be done before hooking up to a patient. The prerequisite guide is the filter replacement for the gas module. Preventive maintenance Maintenance kit, 5,000 hours Only original parts from the manufacturer must be used. Spare parts and maintenance kits can be ordered from your local Maquet representative. When performing this maintenance, a 'Maintenance kit, 5,000 hours' should be used. The following parts shall be replaced and they are included in the 'Maintenance kit, 5,000 hours': A. Filters for the gas modules B. Nozzle units for the gas modules C. Bacteria filter for the inspiratory pressure transducer D. Bacteria filter for the O2 cell Performing the Preventive maintenance: • Disassembling and assembling of the unit is required when replacing parts included in the 'Maintenance kit, 5,000 hours'. If not stated otherwise, refer to chapter 'Disassembling and assembling' for instructions. • The letters A – D in the text below refers to the description of the Maintenance kit above. • Prepare the unit as described in section 'Preparations' above in this chapter Service Manual 7-3