FIXONIC
SellSavedMy Fixonic
Official iFixit API visual reference

Philips Emerson CoughAssist CA Series Disassembly

If your CoughAssist ventilator is acting up and...

iFixit Library Philips Emerson CoughAssist CA Series Moderate 5 - 10 minutes Guide 134057
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

If your CoughAssist ventilator is acting up and not working correctly, the first step to replacing parts is always disassembling the main unit. This guide will allow you to start repairing almost any part within the ventilator.

Step 1

Repair step

  • First, remove the two 16mm pan head screws (1) directly under the front panel of the ventilator. (Page 3).
Step 2

Repair step

  • Next, remove the two 16mm flat head screws (3) from the back panel of the ventilator. (Page 3).
Step 3

Repair step

  • Lift off the top cover (2) of the device. (Page 3).
Step 4

Repair step

  • Now remove the "Inhale flow" control knob (4) and "Inhale pressure" knob (4), by loosening the setscrew on each before pulling off. If you have a CM model, carefully remove the Manual Control Lever (3) by prying it off with the screwdriver. (Page 3).
Step 5

Repair step

  • Unscrew the three 12.7mm plastite screws ( 2) holding the panel assembly to the lower housing. A long Phillips screwdriver is very helpful for access to the three screws. (Page 3)
Step 6

Repair step

  • The front control panel (5) should now be loose. Carefully slide it forwards to remove. (Page 4).