FIXONIC
SellSavedMy Fixonic
Official iFixit API visual reference

Hamilton G5 Expiratory Valve Replacement

For more information on replacing the...

iFixit Library Hamilton G5 Moderate 10 minutes Guide 134065
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

For more information on replacing the expiratory valve, please refer to section 10 of the service manual.

Step 1

Repair step

  • Remove 4 Hex Allen screws (A) from the Rail surrounding the Ventilation Unit.
Step 2

Repair step

  • Remove the Rail (B).
Step 3

Repair step

  • The Standoff Brackets (C) remain with the Main Cover.
Step 4

Repair step

  • Place aside in a safe location.
Step 5

Repair step

  • Lift the Main Cover (A) from the Ventilation Unit using the Standoff Brackets (B) attached to the Main Cover.
Step 6

Repair step

  • Place aside in a safe location.
Step 7

Repair step

  • Disconnect 2 cables from the magnet of the Expiratory Valve.
Step 8

Repair step

  • Remove the Expiratory Valve Cover (A) and Silicon Membrane (A) (if not already removed).
Step 9

Repair step

  • Remove 3 Hex Allen screws (B) M3x8
Step 10

Repair step

  • Remove the Expiratory Valve.