Before you begin

This guide can help you replace the alarm light board (PWA), part M1227419-S), for your GE Carescape R860. If the alarm light is malfunctioning, replacing the PWA board may be necessary. To get access to the PWA board, one must first remove the rear housing and shield of the display unit. For instructions on removal, refer to the Display Unit Disassembly. For additional information regarding the replacement of the alarm light board (PWA), see section 9.5.3 (pg. 251) of the GE CARESCAPE R860 Technical Reference Manual.

Step 1

Repair step

  • Using an opening tool, gently pry the alarm light lens (boxed) from the display unit.
Step 2

Repair step

  • Loosen the seven screws that hold the rear housing using a Phillips #0 screwdriver.
  • Remove the rear housing.
Step 3

Repair step

  • Cut the cable tie holding the two cables at the lower right side using scissors.
Step 4

Repair step

  • Disconnect the touch controller cable.
Step 5

Repair step

  • Disconnect the Keypad harness at the Carrier Board.
Step 6

Repair step

  • Disconnect the encoder harness from the Carrier Board.
Step 7

Repair step

  • Disconnect the speaker harness from the Carrier Board and set the speakers aside.
Step 8

Repair step

  • Disconnect the alarm light harness from the Carrier Board.
Step 9

Repair step

  • Loosen (do not remove) the five screws at the top edge of the shield just enough to raise it and slide the shield away with a Phillips #0 screwdriver.
Step 10

Repair step

  • Loosen the four screws along each side of the shield with a Phillips #0 screwdriver.
Step 11

Repair step

  • Remove the four screws at the bottom edge of the shield with a Phillips #0 screwdriver.
Step 12

Repair step

  • Slide the shield down to disengage the touch controller harness connector.
Step 13

Repair step

  • Pull the PWA board from the slot.
  • Replace the PWA board.
Step 14

Repair step

  • Reconnect the alarm light harness.