Before you begin

Under high usage the blower assembly is under a large amount of stress causing the whole unit to malfunction. Thus, needing to replace the whole blower. For more information on replacing the blower assembly, please refer to page 6-16 in the service manual.

Step 1

Repair step

  • Remove the UI Knob by pulling it straight out.
Step 2

Repair step

  • Remove the top cover by using a Torx T15 to remove the 2 screws on the back of the enclosure.
Step 3

Repair step

  • Remove the Main PCA by removing the wiring harnesses and squeezing the 2 standoffs in the corners of the board.
Step 4

Repair step

  • Remove the Blower Cap by using the T8 Torx screwdriver to remove the 7 screws that secure the Cap to the enclosure.
Step 5

Repair step

  • Lift the Blower Assembly out of the blower housing.
  • Remove the blower Outlet Bellows from the blower assembly.
Step 6

Repair step

  • Place the New outlet bellows onto the blower assembly.
Step 7

Repair step

  • Align the stands located on the bottom of the blower assembly with the holes in the blower housing.