GE Carescape R860 Air Filter Cleaning

In this guide, the user will learn how to...

iFixit Library GE Carescape R860 Easy 5 minutes Guide 134097
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

In this guide, the user will learn how to properly disconnect the hoses to the ventilator in order to do a visual inspection for debris. Steps one through three can be found in the GE Carescape support video titled Gas Inlet, O2, and Air Filter and explain how to properly disconnect the hoses from the machine. Steps four and five are from the GE Carescape Manual on page 5-4 and instruct the user on how to clean the filter if debris is found. For the full video, please visit https://services.gehealthcare.com/gehcst...

Step 1

Repair step

  • Disconnect the pipeline sources from the wall.
  • These are the yellow and green hoses, the air and oxygen hoses, in the image.
Step 2

Repair step

  • Disconnect the oxygen hose from the machine.
  • This also includes the removal of the oxygen fitting.
Step 3

Repair step

  • After the oxygen hose has been disconnected, the air hose will be disconnected as well.
  • The air line may have a compressor hose, like this unit, which will need to be removed before the air fitting can be removed.
  • Once this is done, you can look through both openings with a flashlight for debris. If there is debris, move onto the next step. Otherwise, you should reinstall the hoses to the machine and the wall
Step 4

Repair step

  • Open both shut-off valves.
  • Verify that the inspiratory outlet is open to atmosphere (not plugged).
  • On the VCB screen:
  • Verify that the Air and O2 Flow Valves are energized ().
  • Set the Air DAC counts to 60,000 (fully open).
  • Set the O2 DAC counts to 60,000 (fully open).
Step 5

Repair step

  • After 20 seconds:
  • Close both shut-off valves. Pulse the Air and O2 flow valves 5 times for 5 seconds.
  • Set the Air DAC counts to zero (closed).
  • Set the O2 DAC counts to zero (closed).
  • Note: Closing the shut-off valves before setting the flow DAC counts to zero, bleeds pressure from the system.