Before you begin

Reference page 28-30 from the Service and Repair Instructions linked under Documents on the Main Page.

Step 1

Repair step

  • Remove the covering cap. (pg. 23)
Step 2

Repair step

  • Place the unit upside down on a non-slip surface and remove the six screws. (pg. 23)
Step 3

Repair step

  • Remove the lower part of the housing with a swiveling motion away to the side. Note the SOMNOcomfort series uses a shorter circuit board than the balance series ventilator. (pg. 23)
Step 4

Repair step

  • Remove the connector (X1) from the circuit board.
  • Pull the connecting cable (64) from its mount.
  • Now set aside the lower part of the housing. (pg. 23)
Step 5

Repair step

  • Next, how to close the compartment and reverse the process.
  • Hold the lower part of the housing to the upper part at an angle
  • Attach the connector (X1) to the main circuit board. When doing so the connector’s nub has to point towards the lock. (pg. 24)
Step 6

Repair step

  • Properly align the main circuit board inside the unit.
  • Attach the connecting cable (64) to its mount (pg. 24)
Step 7

Repair step

  • Place the lower part of the housing on the upper part (47). Please make sure the main circuit board (44) is aligned properly, and that the connecting cable (64) is not being squeezed anywhere. (pg. 24)
Step 8

Repair step

  • Now reattach the upper part of the housing by tightening the six screws. (pg. 24)
Step 9

Repair step

  • Insert the covering cap. (pg. 24)
Step 10

Repair step

  • Remove baffle box.
Step 11

Repair step

  • Remove the decoupling tube(50) from the box.
Step 12

Repair step

  • Remove the box(53) in an upwards motion. Leave the Foam bottom (59) attached.
Step 13

Repair step

  • Remove the four screws (69) in the cover of the baffle box (60).
  • Remove the foam cover (58) and the cover of the baffle box (60).
Step 14

Repair step

  • Remove the foam bottom and separate the two parts of the baffle box.
Step 15

Repair step

  • Pull the decoupling tube(55) out of the motor frame’s borehole7.
Step 16

Repair step

  • Loosen the motor mount(54). Cut the mount using scissors or scalpel.
  • Remove the Fan.
Step 17

Repair step

  • Place three new motor mounts(54) into the box.
Step 18

Repair step

  • Insert the fan into the motor frame and attach it using the motor mounts(54). Pull the motor mounts through the boreholes of the fan to reach the surrounding notch.
Step 19

Repair step

  • Apply a small amount of 70% Isopropanol solution on the decoupling tube’s(55) circumference and attach it to the box. The tube’s notch must completely lock into the bore hole rim.
Step 20

Repair step

  • Reconnect the two parts of the box. Pay attention to the proper alignment of the foam part (61) and the correct cable lead. The cable must be placed in the appropriate cut-out.
Step 21

Repair step

  • Place the cover on the box and insert the screws.
Step 22

Repair step

  • Insert Baffle Box. Place foam 58on the box, soft side up. You must be able to see the screw heads of the box through the holes in the foam
Step 23

Repair step

  • Insert box(53) and foam (58) into the housing(47). Make sure the foam is spread out evenly.
Step 24

Repair step

  • Pull the protective backing off the adhesive tape and stick on foam(59). Use the crosses as a positioning aid.
Step 25

Repair step

  • Apply a small amount of 70% Isopropanol solution on the decoupling tube’s (50) circumference and attach it to the box. The tube’s notch must completely lock into the bore hole rim.