How to Manually Reset the Touch Screen (Hard Reset)

If the touch screen still does not respond as...

iFixit Library Puritan Bennett 980 Ventilator Easy 2 minutes Guide 134154
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

If the touch screen still does not respond as expected after following the guide for a soft reset of the display, the operator should provide an alternate source of ventilation to the patient before performing the hard reset steps in this guide. Further information regarding the reset of the touch screen can be found on the Puritan Bennett 980 support page in the video titled “Unresponsive or Blank GUI Reset”.

Step 1

Repair step

  • Before performing the following steps, make sure to safely transfer the patient to another ventilation system.
  • Power the device off and back on using the power switch on the front of the device.
Step 2

Repair step

  • If appropriate, the patient can be transferred back to the device.
  • If the patient is going to be transferred back to the ventilator, first press the “Same Patient” button on the display.
  • Press the “Start” button to resume the patient profile.