Newport e500 Pneumatic Box Disassembly

The pneumatic box is a very important removal...

iFixit Library Newport e500 Difficult 15 minutes Guide 134160
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

The pneumatic box is a very important removal step and requires the user to be very careful when removing. It is important to note that the guide removal begins on page 44 of the service manual. When reinstalling, the user must read the reinstalling instructions located on page 45 of the service manual.

Step 1

Repair step

  • Place device so that the top cover is facing you.
  • Open the top cover.
Step 2

Repair step

  • Remove the four hex nuts and lock washers (two at each hinge) that secure the cover to the hinges on the left side of the pneumatic box.
Step 3

Repair step

  • Remove the cover and set aside. Retain hardware for re-assembly.
Step 4

Repair step

  • Carefully place the GDU on its right side with the front of the GDU facing you.
  • Remove the three screws and washers at the top edge of the left side panel.
Step 5

Repair step

  • Remove the two screws and washers from the underside of the left side panel.
Step 6

Repair step

  • Remove the left side panel. Retain hardware for re-assembly.
Step 7

Repair step

  • Carefully place the GDU on its left side.
Step 8

Repair step

  • Remove the four screws and washers at the top edge of the right side panel.
Step 9

Repair step

  • Remove the two screws and washers from the underside of the right side panel.
Step 10

Repair step

  • Remove the right side panel. Retain hardware for re-assembly.
Step 11

Repair step

  • Remove the three screws and washers that secure the flow sensor cover to the pneumatic box assembly. Retain the hardware for re-assembly.
Step 12

Repair step

  • Remove the flow sensor cover.
Step 13

Repair step

  • Pull up and remove the flexible rubber exhalation outlet boot (push fit).
Step 14

Repair step

  • Disconnect the tubes from both sides of the flow sensor and the two tubes from the exhalation valve.
Step 15

Repair step

  • Pull the exhalation module straight up to remove it from the GDU.
Step 16

Repair step

  • Remove the two screws and washers (back corners), one screw and washer (exhalation module support bracket), and two standoffs and washers (front corners) that secure the pneumatic (upper) box assembly to the electronic box assembly. Lift the Pneumatics Box and carefully tilt to one side.
Step 17

Repair step

  • Remove cables from connectors J1, J2, J3, J4, J5, and J6 on the main PCB (4 connectors are not shown due to being blocked by the side panel). Note the position of these connectors for re-assembly.
Step 18

Repair step

  • If not already detached, remove the heater cables from J5 and J6 on the DC-DC power supply PCB.