Newport e500 Power Supply Disassembly

This step of removing the power supply needs to...

iFixit Library Newport e500 Moderate 5 minutes Guide 134161
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

This step of removing the power supply needs to be done carefully. It is important to note that in no instance are you able to touch the transformer on the power supply. This could lead to permanent damage. For more information go to page 75 of the service manual.

Step 1

Repair step

  • Place device so that the top cover is facing you.
  • Open the top cover.
Step 2

Repair step

  • Remove the four hex nuts and lock washers (two at each hinge) that secure the cover to the hinges on the left side of the pneumatic box.
Step 3

Repair step

  • Remove the cover and set aside. Retain hardware for re-assembly.
Step 4

Repair step

  • Carefully place the GDU on its right side with the front of the GDU facing you.
  • Remove the three screws and washers at the top edge of the left side panel.
Step 5

Repair step

  • Remove the two screws and washers from the underside of the left side panel.
Step 6

Repair step

  • Remove the left side panel. Retain hardware for re-assembly.
Step 7

Repair step

  • Carefully place the GDU on its left side.
Step 8

Repair step

  • Remove the four screws and washers at the top edge of the right side panel.
Step 9

Repair step

  • Remove the two screws and washers from the underside of the right side panel.
Step 10

Repair step

  • Remove the right side panel. Retain hardware for re-assembly.
Step 11

Repair step

  • Remove the three screws and washers that secure the flow sensor cover to the pneumatic box assembly. Retain the hardware for re-assembly.
Step 12

Repair step

  • Remove the flow sensor cover.
Step 13

Repair step

  • Remove the two screws and washers (back corners), one screw and washer (exhalation module support bracket), and two standoffs and washers (front corners) that secure the pneumatic (upper) box assembly to the electronic box assembly. Lift the Pneumatics Box and carefully tilt to one side.
Step 14

Repair step

  • Remove cables from connectors J1, J2, J3, J4, J5, and J6 on the main PCB (4 connectors are not shown due to being blocked by the side panel). Note the position of these connectors for re-assembly.
Step 15

Repair step

  • If not already detached, remove the heater cables from J5 and J6 on the DC-DC power supply PCB.
Step 16

Repair step

  • From the underside of the electronics box assembly, remove the four screws and lock washers that secure the power supply to the box assembly.
Step 17

Repair step

  • Disconnect cables from connectors J1 and J2 on the power supply.
Step 18

Repair step

  • Lift the power supply out of the electronics box assembly.