Puritan Bennett 560 Internal Battery Replacement

This guide will explain how to remove and repla...

iFixit Library Puritan Bennett 560 Easy 3 minutes Guide 134163
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

This guide will explain how to remove and replace the internal battery in the Puritan Bennett 560 ventilator. Warnings: Ensure the ventilator is powered off and AC power is disconnectedDo not try to open the battery itself under any circumstancesCover the battery terminals with a non-conductive material when not in use to prevent short-circuiting. The repair information is also available on pages 143-144 (Section 5, Pages 13-14) of the Service Manual that is linked in the documents section.

Step 1

Repair step

  • REMOVAL
  • Locate the battery cover on the rear of the ventilator.
Step 2

Repair step

  • Using the T10 Torx screwdriver, remove the three 8mm screws on the battery cover.
  • Two screws are located on the rear of the ventilator, and one on the bottom.
  • Slide the battery cover out of the ventilator.
  • Remove the battery from the battery cover.
Step 3

Repair step

  • REPLACEMENT
  • Install the battery into cover, as shown in figure.
  • Make sure the contacts on the battery line up with those in the ventilator.
  • Slide the battery and cover together into the housing.
  • While applying pressure to the cover, install the two rear screws.
  • Install the screw at the bottom of the cover.