FIXONIC
SellSavedMy Fixonic
Official iFixit API visual reference

Fisher Paykel Neopuff Manometer Replacement

Please refer to page 7 of the device’s...

iFixit Library Fisher Paykel Neopuff Moderate 15 minutes Guide 134187
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

Please refer to page 7 of the device’s Technical Manual for source information.

Step 1

Repair step

  • Remove the back cover, fixed by four screws.
Step 2

Repair step

  • Disconnect the tube from the manometer.
Step 3

Repair step

  • Remove the manometer by unscrewing the two retaining nuts.
Step 4

Repair step

  • Fit the new manometer into the front panel.
Step 5

Repair step

  • Tighten the retaining nuts.
Step 6

Repair step

  • Reconnect the manometer tube.
Step 7

Repair step

  • Refit the front panel to the back cover with the four screws.
Step 8

Repair step

  • Carry out the manometer performance test. It is recommended to record the lot number from the box label of the replacement manometer on the maintenance checklist.