Philips Respironics V60 Touchscreen Replacement

The Touchscreen is the main command function of...

iFixit Library Philips Respironics V60 Easy 10 minutes Guide 134194
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

The Touchscreen is the main command function of the Ventilator and is the prime source to navigate the different functions of the Ventilator. PLEASE NOTE: This repair guide was developed by the iFixit team based on Philips's own service manual. Neither iFixit nor this repair guide is endorsed by or affiliated with Philips. Re-Calibration of the touchscreen is discussed in: chapter 5.3 “Diagnostic Mode and Troubleshooting - Touch Screen Calibration” of service manualAppendix F “Diagnostic mode - Touchscreen calibration” of user manual Assembly of touchscreen is discussed in: chapter 7.32 “Component Removal/Installation - Front Bezel, Touch Screen” of service manual

Step 1

Repair step

  • From the Diagnostics Menu, select Touch Screen Calibration. The Calibrate Touch Screen screen is displayed.
  • Follow the steps shown. Press on the middle of each target with a blunt, narrow object.
  • NOTE: If the Touch Screen Calibration button does not respond, press the Accept button on the navigation ring to begin
  • If the calibration is not successful, have the ventilator serviced.
Step 2

Repair step

  • Press and release the ON/Shutdown key, opening the Shutdown window in the screen.
  • Select Ventilator Shutdown, which will shut down the ventilator.
  • Improper shutdown may cause a Power has been restored message the next time the ventilator is turned on.
Step 3

Repair step

  • Remove the screws from the end cap and remove the end cap.
  • Remove the screws from the rear bezel.
  • Separate the front and rear bezel.
Step 4

Repair step

  • Disconnect these cables from the UI PCBA: • Switch PCBA cable • Nav-ring cable • Power switch overlay cable • Touch screen cable
  • Remove the switch PCBA.
  • Remove the screw that attaches the cable clamp to the LCD tray.
  • Remove the screws and washers that attach the LCD tray to the front bezel, then remove the LCD tray.
  • Remove the touch screen from the front bezel.