Philips Respironics V60 Alarm Speaker Replacement

The alarm speakers are to sound and notify if...

iFixit Library Philips Respironics V60 Moderate 30 minutes Guide 134197
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

The alarm speakers are to sound and notify if an issue or emergency is to occur to the patient currently in use of the ventilator of is a problem occurs within the ventilator itself. PLEASE NOTE: This repair guide was developed by the iFixit team based on Philips's own service manual. Neither iFixit nor this repair guide is endorsed by or affiliated with Philips Replacing alarm speaker is discussed in: chapter 7.14 “Component Removal/Installation - Speakers” of service manualchapter 7.10 “Component Removal/Installation - Front Panel” of service manualchapter 7.9 “Component Removal/Installation - Separating the UI from the Base” of service manualchapter 7.8“Component Removal/Installation - Motor Controller (MC) PCBA” of service manualchapter 7.4 “Component Removal/Installation - Top Cover” of service manual

Step 1

Repair step

  • step 1 Remove MC PCBA
  • Remove top cover (top cover guide here)
  • Remove the screws and wing nuts that attach the MC PCBA to the PCBA brackets.
  • Loosen the captive screws that attach the capacitor bracket to the base.
  • Disconnect cables: • Blower cable • Fan cable • DA-MC ribbon cable
  • Gently pull MC PCBA up from its connection on the CPU PCBA, then disconnect speaker #1 cable.
  • Remove PCBA brackets. To reinstall, follow the steps above in reverse order
Step 2

Repair step

  • step 2 separate the UI from the base
  • Remove MC PCBA (step 1)
  • Disconnect cables: • LCD cable from PM PCBA. • LCD cable ground wires from the ground terminal. • LCD tray ground wires from the ground terminal. • UI cable (ribbon cable) from the PM PCBA.
  • Loosen the captive screw that attaches the UI retainer to the UI.
  • Slide retainer up, then remove UI from base.
Step 3

Repair step

  • step 3 remove the front panel
  • Separate UI from base (step 2)
  • Remove hose clamp and boot that attach blower to the gas outlet.
  • Remove the tubing from the barbed fittings inside the front panel.
  • Disconnect gas outlet and proximal port ground wires from the base.
  • Remove the screws that attach the front panel to the base and side walls.
  • NOTE: When reinstalling tubing, ensure that: • Longer (9-in.) tube is connected between manifold barb “M” (machine) to barbed gas outlet fitting. • Shorter (8-in.) tube is connected between manifold barb “P” (proximal) and barbed proximal pressure port.
Step 4

Repair step

  • step 4 remove the alarm speakers
  • Remove front panel (step 3)
  • Disconnect speaker #2 from the power management (PM) PCBA. (Speaker #1 is disconnected when the MC PCBA is disconnected.)
  • Remove the screw and washer that attaches each speaker to the base.
  • Insert new speaker in place of old one
  • Follow step 4 in reverse order to install speaker.
Step 5

Repair step

  • step 5 reinstalling
  • Follow step 3 in reverse order to reinstall the front panel.
  • Follow step 2 in reverse order to reinstall the UI to the base.
  • Follow step 1 in reverse order to reinstall the MC PCBA.
  • Follow the Top Cover Guide in reverse order to reinstall.