How to Adjust the Max Pressure Relief on the Fisher Paykel Neopuff

Please refer to page 7 of the device’s...

iFixit Library Fisher Paykel Neopuff Moderate 15 minutes Guide 134199
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

Please refer to page 7 of the device’s Technical Manual for source information.

Step 1

Repair step

  • Adjust the gas flow to 10 L/min.
Step 2

Repair step

  • Close the Peak inspiratory Pressure valve by turning the knob fully clockwise.
Step 3

Repair step

  • Adjust the Max Pressure Relief knob clockwise or counterclockwise until manometer reads 40 cmH20
Step 4

Repair step

  • Turn the Peak Inspiratory Pressure knob counterclockwise so the manometer reads 20 cmH20 and shut off the gas flow.