Newport e360 Exhalation Flow Sensor Replacement

Replacing the Exhalation Flow Sensor should be...

iFixit Library Newport e360 Easy 10 minutes Guide 134201
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

Replacing the Exhalation Flow Sensor should be relatively quick and easy. There are no tools needed in this replacement. Make sure to give yourself time to thoroughly read through the instructions. For more information on replacing the exhalation flow sensor, please refer to page 4-2 in the service manual.

Step 1

Repair step

  • Turn off ventilator before starting this repair.
  • Shut off or disconnect the air and oxygen gas supplies before starting this repair.
Step 2

Repair step

  • To avoid damaging equipment, always use standard electrostatic discharge (ESD) precautions
  • Put on an ESD wrist strap or ESD gloves
Step 3

Repair step

  • Open the Front Panel Door on the lower left front of the ventilator to expose Exhalation Valve EXH2105A and Flow Sensor FLS2101P.
Step 4

Repair step

  • Remove the Exhalation Valve EXH2105A by releasing the retaining latch.
Step 5

Repair step

  • Disconnect the Flow Sensor Cable CBL2123P from the plastic body of the Flow Sensor FLS2101P by pulling cable straight up. Do not twist.
Step 6

Repair step

  • With a twisting motion, pull the plastic Flow Sensor FLS2101P away from the outlet of the Exhalation Valve EXH2105A.