Newport e360 Top Cover CVR2101M Replacement

The top cover is used as the housing unit for...

iFixit Library Newport e360 Easy 5 - 10 minutes Guide 134203
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

The top cover is used as the housing unit for the wiring and electronics in the inside of the ventilator. In order to access the inside of the ventilator the top cover needs to be removed. The following steps are not difficult but be sure to thoroughly read all the steps. For more information on removing the top cover, please refer to page 4-3 in the service manual.

Step 1

Repair step

  • Turn off the Ventilator before starting this repair.
  • Shut off or disconnect he air and oxygen gas supplies.
Step 2

Repair step

  • For your safety put on an electrostatic discharge (ESD) wrist strap.
  • To adviod damage to the ventilator always used standard ESD precaustions while performing any service.
Step 3

Repair step

  • Rotate the device so that the back face is facing you.
  • The back face will look like the picutre.
Step 4

Repair step

  • Locate and remove the 6 screws (Philips size 1) that secure the cover.
  • The picture locates 3 screws on the left side. The other 3 screws are found at the same location on the right side.
Step 5

Repair step

  • Grab the handles on the side and pull the cover away from the rest of the device body.