FIXONIC
SellSavedMy Fixonic
Official iFixit API visual reference

Hamilton G5 Ventilation Unit Motherboard Replacement

For more information on replacing the...

iFixit Library Hamilton G5 Easy 10 minutes Guide 134207
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

For more information on replacing the motherboard, please refer to section 10 in the service manual.

Step 1

Repair step

  • Remove 4 Hex Allen screws (A) from the Rail surrounding the Ventilation Unit.
Step 2

Repair step

  • Remove the Rail (B).
Step 3

Repair step

  • The Standoff Brackets (C) remain with the Main Cover.
Step 4

Repair step

  • Place aside in a safe location.
Step 5

Repair step

  • Lift the Main Cover (A) from the Ventilation Unit using the Standoff Brackets (B) attached to the Main Cover.
Step 6

Repair step

  • Place aside in a safe location.
Step 7

Repair step

  • First, we need to remove the ESM module. It is connected to 2 connector sockets. Disconnect and remove.
Step 8

Repair step

  • Take out the Ventilation Real-time Controller (VRC) SW chip
Step 9

Repair step

  • Install the ESM Module, by connecting the connectors (A), and the VRC SW chip on the new board.
Step 10

Repair step

  • Disconnect all cables and connections from the board to the unit.
Step 11

Repair step

  • Remove the three Hex screws (B).
Step 12

Repair step

  • Move the motherboard out then up away from the ventilation unit (C).