Cardinal Health Vela Ventilator Turbine Filter P/N 10365 Replacement

A failure on the "Delivered Volume Test" can...

iFixit Library Cardinal Health Vela Ventilator Difficult 40 minutes Guide 134218
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

A failure on the "Delivered Volume Test" can signify dirty filters needing replacement. Knowing how to replace these is vital to the Ventilator's function. This guide will run through all the necessary steps including the removal of the: -Power cable -Top Panel -Battery Tray -Left Panel -Blender Assembly And finally the Turbine/Muffler Assembly itself

Step 1

Repair step

  • Use a screwdriver to remove the two Phillips pan-head screws at the top of the plug guard.
Step 2

Repair step

  • Remove the plug guard and unplug the ventilator.
Step 3

Repair step

  • Remove the four Phillips #0 screws on the back side of the ventilator.
Step 4

Repair step

  • Remove all six small side panel screws along the bottom of each side of the top cover with the same screwdriver.
Step 5

Repair step

  • Slide the top cover towards the back side of the ventilator to remove it.
Step 6

Repair step

  • Use your Phillips screwdriver to unscrew the 4 Phillips pan head screws positioned on each side of the battery tray.
Step 7

Repair step

  • Lift the battery tray out of the ventilator.
  • Caution: The wires connecting the batteries to the ventilator will still be connected so lift the battery tray carefully.
Step 8

Repair step

  • Disconnect the batteries from the ventilator by pulling apart the Battery Connector.
Step 9

Repair step

  • Remove the two Phillips countersink at the rear of the panel.
Step 10

Repair step

  • Remove the Phillips countersink in the upper front of the panel.
Step 11

Repair step

  • The panel is now free and can be lifted from the device.
Step 12

Repair step

  • To install the left panel follow these steps in reverse order.
Step 13

Repair step

  • If the rear panel is installed, remove the high and low pressure oxygen fittings from the rear panel using your 3/4" wrench. NOTE: the high pressure fitting is located above the low pressure fitting
Step 14

Repair step

  • If the rear panel is installed, disconnect the 40-pin ribbon cable from the SVGA out PCB.
Step 15

Repair step

  • If the front panel is installed, disconnect the green oxygen nebulizer tube.
Step 16

Repair step

  • Disconnect the:
  • -J301, 12-Pin Connector to J301 on the main PCB
  • From the blender PCB
Step 17

Repair step

  • Remove the bottom strip ambient air inlet filter by pinching and pulling it out.
Step 18

Repair step

  • Remove the three Phillips pan-head screws on the right side of the rear panel.
Step 19

Repair step

  • The assembly should now be free to remove. As you remove the unit use your needle-nose pliers to disconnect the oxygen diffuser tube as you remove the blender assembly.
Step 20

Repair step

  • If the exhalation valve assembly is installed, disconnect:
  • J3 [2-pin (blue connection)]
  • To the oxygen inlet solenoid on the blender assembly.
  • J4 [2-pin (red connection)]
  • To the Nebulizer solenoid assembly on upper rear of blender.
Step 21

Repair step

  • For Instillation follow steps in reverse order.
Step 22

Repair step

  • Disconnect the main wire harness connector at P2 on the Turbine Driver PCBA
Step 23

Repair step

  • Remove the 5 Phillips pan-head mounting screws.
Step 24

Repair step

  • Disconnect the gray silicone intake and output elbows by gently pulling them out from the muffler tubes.
Step 25

Repair step

  • The Turbine/Muffler Assembly should now be free to lift off the manifold base.
  • Gently pull up on the elbows to remove
Step 26

Repair step

  • To Install P/N 16350 "Turbine/Muffler Assembly" repeat the steps in reverse order.
Step 27

Repair step

  • Remove the six Phillips pan-head screws on the cover of the filter
Step 28

Repair step

  • The cover should be free now, remove it as well as the two "O"-rings
Step 29

Repair step

  • Use your needle-nose pliers to remove the left and right filter assemblies. The right filter should have a "filter sock" on it.
Step 30

Repair step

  • To Install follow the steps in reverse order. If you are replacing the P/N 10365 muffler/filter assemblies then, make sure to place the muffler (output) in the left tube and the sock filter (input) in the right tube.