FIXONIC
SellSavedMy Fixonic
Official iFixit API visual reference

Bear 1000 Power Entry Module Replacement

This guide will walk you through the process of...

iFixit Library Bear 1000 Moderate 25 - 35 minutes Guide 134221
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

This guide will walk you through the process of replacing the Power entry Module in a Bear 1000 Ventilator. Found on page 88 in the full service manual.

Step 1

Repair step

  • Remove the ribbon cable at the connection on the oxygen blender control board.
Step 2

Repair step

  • Mark the plastic hoses attached to the oxygen blender (so you can put them back on the same way) and then remove.
Step 3

Repair step

  • The air/oxygen blender has two screws close to the control board. Remove them and remove the unit.
Step 4

Repair step

  • Remove the pneumatic and electrical connections labeled.
Step 5

Repair step

  • Remove the water trap bowls on the rear of the unit.
Step 6

Repair step

  • Remove the four screws and remove the manifold.
Step 7

Repair step

  • Remove the two screws holding the power supply to the manifold.
Step 8

Repair step

  • Mark or label the 6 faston connections as to position and remove
Step 9

Repair step

  • Mark or label the two wires going to the power switch and remove at the power switch end
Step 10

Repair step

  • Using an allen wrench, remove the power cord retaining bracket.
Step 11

Repair step

  • Using an allen wrench, remove the mounting screwa and nuts from the Power Entry Module.
Step 12

Repair step

  • After removing the heat shrink tubing, unsolder the two lower wires from the Power Entry Module, noting locations.
Step 13

Repair step

  • Solder the wires to the replacement module, covering the joints with heat shrink tubing.
Step 14

Repair step

  • Transfer the voltage selector drum from the original module to the replacement and reinstall.