Guide overview

Before you begin

This guide will show you the process of removing the circuit board from the Philips Respironics System One REMstar ventilator. The circuit board controls all of the operations of the device, including critical pressure and volume managements, so it’s key that the board is functioning correctly if the device is in use on a patient. Be sure to run thorough diagnostics tests on the device after a replacement has been made to ensure the device is functioning as intended.

Step 1

Repair step

  • Turn the device upside down so that the rubber feet are facing you.
  • Use a Torx TR15 screwdriver to remove the two 46.5mm screws located on the back.
  • Gently separate the panel so the wires are not damaged.
Step 2

Repair step

  • Using a Torx TR10 screwdriver, remove the seven 13.5 mm screws that are securing the inner casing.
  • Lift out the inner casing.
Step 3

Repair step

  • Remove the motor from the bottom portion of the casing and set it aside.
Step 4

Repair step

  • Remove the side panel by lifting it up out of the base.
  • Remove the grouped wire and wire connector.
Step 5

Repair step

  • Remove the circular piece by popping it off to separate the circuit board from the top cover.
  • No tools should be necessary to complete this action.
Step 6

Repair step

  • Separate the green, red, and white wire connectors from their connections on the circuit board.