Repairing the Sanitas SBM67 blood pressure monitor control unit button

This guide shows the repair of a broken trace...

iFixit Library Blutdruckmessgerät Sanitas SBM67 Moderate 1 - 2 hours Guide 199501
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

This guide shows the repair of a broken trace on the SET button of a Sanitas SBM67. In my device, two cracks in the control board caused the failure. Other buttons may be affected by a similar defect.

Step 1

Preparing the device

  • With this Sanitas SBM67, the SET button is no longer responding.
  • Turn the device over and remove the batteries. Then remove the rubber feet at the bottom.
  • Use a Phillips screwdriver to remove the four screws: two in the battery compartment and two under the bottom rubber feet.
Step 2

Opening the housing

  • Carefully release the six clips around the display using an opening tool. Remove the upper part of the housing.
Step 3

Repairing the circuit board

  • Remove the screws on the left and right to release the PCB with the five switch contacts.
  • In the device shown, two cracks in the control board have broken the ground connection for the SET button.
  • Bridge the broken circuit path by soldering an insulated wire. Determine the two solder points belonging to the broken ground path beforehand; do not solder as long as this assignment is unclear.
  • Observe the soldering basics and safety guidelines.
  • With the batteries removed, check the connection and inspect the solder joints for unintended bridges.