Failure 1: SACI DP35CM Industrial Voltmeter 24/7 use Does not turn on, failure of 2 470 uF/16V capacitors

Attributed iFixit repair guide for Voltimetro industrial SACI DP35CM.

iFixit Library Voltimetro industrial SACI DP35CM Moderate 1 hour Guide 224823
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Step 1

Repair step

  • SACI industrial voltmeter, a Spanish brand; it has been in use for quite a few years and broke down by turning off. It was replaced with a new one from another manufacturer, but why not take a look and try to repair it as a challenge?
Step 2

Repair step

  • Disassembling it, I see that it is made up of 2 PCB boards stacked one on top of the other, the smaller board being the multi-voltage power input (very typical in the industrial field) and that the output is a symmetrical power supply of +12V, 0, -12V, so this voltage can be very easily simulated with 2 dead batteries and checked.
  • If the problem is only with the power input board or has also affected the measurement board, with the 2 dead batteries connected, the board turns on, so the problem is only with the power input board.
  • I see that it turns on with 12V and 0, but it needs the -12V to be able to stabilize the display so that it stays at 000.0V
Step 3

Repair step

  • Example: by removing the -12V voltage, the display does not stay at 000.0V
  • Photo 3: view from where I have taken the wiring to power it with external batteries to test the board.
Step 4

Repair step

  • View of the power input "backpack" board, which is causing the failure.
Step 5

Repair step

  • I desolder the board to perform measurements and be able to remove components from it. Measuring the capacitors directly on the PCB, 2 of them provide a measurement much lower than what they should be measuring; I remove them from the PCB to measure them outside.
Step 6

Repair step

  • The two capacitors measured off the PCB, rated at 470 µF/16 V, measure approximately 60.7 µF; they are completely low on capacitance.
  • I replace both and test the PCB by applying 230 V AC input voltage.
Step 7

Repair step

  • View of the 2 replaced capacitors; since I am using recycled material from other boards, I soldered one of them with 2 pieces of wire because its terminals were too short.
Step 8

Repair step

  • With the circuit powered at 230 V, it now has a +12 V, 0, -12 V voltage output.
Step 9

Repair step

  • Measured between +12 and -12 we have the nearly 24 V difference, so the backpack board can be soldered back onto the main board.
Step 10

Repair step

  • Mounting the backpack PCB onto the main PCB.
Step 11

Repair step

  • View of the voltmeter once repaired; having replaced the two 470 µF / 16 V capacitors, we once again have an industrial panel voltmeter for 24/7 use. This one was replaced by a new one, so now I have a functional spare.