Zoll AED Plus Battery Replacement

Here's how to replace the batteries on your...

iFixit Library Zoll AED Plus Easy 45 seconds - 5 minutes Guide 240331
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

Here's how to replace the batteries on your ZOLL AED Plus!

Step 1

Open the Battery Cover

  • Open the Battery Cover: Turn the AED face down. Insert a coin or tool into the two slots at the bottom of the back cover to release the locking latches, then lift the cover off.
Step 2

Remove Old Batteries

  • Remove Old Batteries: Take out all 10 old batteries at once and discard them properly. Never mix old and new batteries.
Step 3

Insert New Batteries

  • Insert New Batteries: Place the new batteries into the well, paying heed to the positive and negaive ends.
Step 4

Press the Reset Button

  • Press the Reset Button: Once the final central batteries are in, the unit will power on and prompt “If new batteries press button”. Press and hold the small blue or green reset button located above the center batteries within 15 seconds to reset the internal power gauge.
Step 5

Secure the Cover

  • Secure the Cover: Slide the battery compartment cover back into its grooves and press down until it snaps firmly into place.
Step 6

Verify Status

  • Verify Status: Flip the unit right-side up and check the status indicator window on the front. Confirm that a green check mark appears and the unit is not beeping. If it is, check the pads, and if they are ok, your AED may need authorised service.