FIXONIC
SellSavedMy Fixonic
Official iFixit API visual reference

Cen-Tech Seven Function Digital Multimeter Ball Bearings Replacement

If the dial cannot spin then it is likely that...

iFixit Library Cen-Tech Seven Function Digital Multimeter Moderate 10 - 20 minutes Guide 32071
Open the authorised source
Licensed iFixit integration

This attributed iFixit guide is displayed through Fixonic's recorded commercial API permission. iFixit remains the independent source and author of the procedure.

Guide overview

Before you begin

If the dial cannot spin then it is likely that the ball bearings are missing or not properly lubricated.

Step 1

Position Device

  • Place the meter face down, and the back panel and screws can be seen.
Step 2

Remove Back Panel Screws

  • Unscrew the two 9mm screws on the back panel using a #0 Phillips screwdriver.
Step 3

Remove Back Panel

  • Remove the back panel of the device by lifting upward.
  • Remove the battery from the meter before continuing.
Step 4

Unscrew Motherboard

  • Remove the four 7mm Phillips screws holding the circuit board in place.
Step 5

Remove Motherboard

  • Gently lift the circuit board away from the device casing.
  • Remove the battery after removing the back cover to reduce the chances of short-circuiting the circuit board.
Step 6

Locate Dial

  • The dial is located in the central part of the device.
Step 7

Remove Dial

  • Carefully remove dial from device by lifting upward.
  • Clean existing dial or replace if necessary.
  • Take care not to lose the small ball bearings.
Step 8

Locate and Remove Ball Bearings

  • Locate ball bearings. Carefully remove.
  • If both ball bearings are still present, try lubricating ball bearings with multi-use lubricant.
  • If one or both ball bearings are missing, simply replace.