Before you begin

This guide will show you how to replace the dead batteries of a Dexcom G4 "slim" transmitter. The Dexcom G4 features a molded plastic body encasing two coin-cell batteries. Replacing the batteries will require grinding away the plastic over the cells, cutting them out, and refilling the excavated plastic with epoxy. Be very careful not to damage the battery cells when grinding!

Step 1

Marking the positions of the batteries

  • Take a marker and paint two circles at this position.
Step 2

Slowly grind down to the batteries

  • Use the 2-3mm cutter and grind slowly down to the batteries.
  • Be careful not to grind away the contact plates.
Step 3

Move up both contact plates with a screwdriver

  • Lever both contact plates carefully with the screwdriver.
Step 4

Cutting around the batteries

  • Use a 0.5mm cutter and grind around the batteries.
  • Be careful not to grind too deep! This process is only for loosen them and make space for the screwdriver.
Step 5

Lever the first battery with the screwdriver

  • The lever part is easy but be careful with the lower contact plates. They are sticking at the (-)pole.
  • Use the edge of the screwdriver and lever it carefully left-right-left ... till it's detached. Then push the contact plate down.
Step 6

The same with the second battery

  • Insert wisdom here.
Step 7

Assembly.

  • The original batteries are Maxell SR1120W (391) with a capacity of 55mAh. I suggest that you use the same ones because of the profile form inside of the transmitter.
  • If you prefer to use another brand, I can suggest Energizer 391 (60mAh) and Renata 391 (50mAh). First you have to scratch the contact plates with the screwdriver.
Step 8

Attach the new batteries

  • Apply some superglue at three points in the battery hole sidely.
  • Then insert the new battery and push it down for 2 minutes.
  • Do the same with the second battery.
  • After this the (-)poles of the batteries should have contact with the lower plate.
Step 9

Attach upper contact plates

  • First you have to scratch the upper sides of the batteries with the screwdriver.
  • Then add a small amount of superglue under the plate and push it down for 2 minutes. Same with the other battery.
  • Do the same with the other battery.
Step 10

Check out for voltage

  • At this point, the transmitter should work again. Let's check it out and take the multimeter.
  • Set it to 2000mV and add the measuring pens on the sensor contacts. It should show 019.
  • If yes ... congratulations! You can move on to the next step.
  • If not, there's a contact problem. First, check the upper contact plates. Set the multimeter to continuity.
  • Touch one pen on the plate and the other on the battery. 000 should be displayed.
  • If not, you have found the problem. You have to remove it, clean it, and re-glue it.
  • If both plates have continuity, you have to check out the lower contacts and remove the batteries again. Maybe pull up the lower contact a little bit before attaching the battery.
Step 11

Prepare 2K Epoxy Glue

  • Slip on some latex gloves and add some 2K-Epoxy glue on a piece of plastic, glass or something else.
  • Mix it about 2 minutes with a toothpick.
Step 12

Sealing

  • Use the toothpick to add the epoxy glue on the batteries and seal the grinded holes completely.
  • After 2Std, it's 80% dried, and you can insert it into your sensor pod.
  • Wait 24 hours before showering.
  • All done! Have fun with your new transmitter. It should now last for about a year. After that ... you now know what to do. Cheers :)